HIOKI 3174 AC Automatic Insulation/ Withstanding HiTester
Contact check and full remote control
Continuous testing of insulation (500/1000 V) and withstand voltage (100 VA transformer capacity). Full remote operation when used in combination with the Safety Test Data Management Software 9267. Save up to 8 test settings each for the withstanding and insulation testing modes. Precise test voltage without power voltage dependency is generated using the PWM method.
Basic specifications (Accuracy guaranteed for 1 year)
[Withstanding test]
Testing voltage : 0.2V AC to 5.00kV AC
Voltage setting method : Digital setting, Setting resolution: 0.01kV
Waveform/ Frequency : Sine wave (Distortion ratio 5% or less at no load), 50/60Hz selectable
Current measurement range : 0.01mA to 20.0mA, True RMS rectified (digital display)
Measurement range : 10mA (0.01mA resolution), 100mA (0.1mA resolution)
Voltage meter : Accuracy: ±1.5% rdg. (1000 V or more), ±15V (less than 1000V), True RMS rectified
Judgment function : Window comparator method (Digital setting)
[Insulation test]
Testing voltage : 500V DC, 1000V DC
Rated testing current : 1 to 1.2mA, Shorted current: 4 to 5mA (at 500V), 2 to 3mA (at 1000V)
Measurement range, Accuracy : 0.5MO to 999MO (at 500V), and 1MO to 999MO (at 1000V) : ±4% rdg., 1000MO to 2000MO : ±8% rdg.
Judgment function : Window comparator method (Digital setting)
[Timer]
Measurement range, Accuracy : 0.3 to 999s, and 0.3 to 99.9s : ±50ms, 100 to 999s : 0.5s
Delay/Ramp timer : 0.1 to 99.9s: ±50ms, 0.1s resolution
Functions : Saving eight testing conditions, hold, buzzer, contact check
Moniter function : Output voltage, detected current, insulation resistance, Refresh rate: 4times/s
Power consumption : 100 to 240V AC, (50/60Hz), 200VA max.
Dimensions, mass : 320mm (12.60in) W × 155mm (6.10in) H × 395mm (15.55in) D, 15kg (529.1oz)
Supplied accessories : H.V. TEST LEAD 9615 (Red, black) × 1, Power cord × 1, Operation manual × 1