HIOKI 3541 Resistance HiTester
Wide measurement range & high-resolution, Optimized for Automated Systems
Wide Measurement Range, 0.1 µO(20 mO range) to 110.000 MO. High Speed & High Precision Measurements, As fast as 0.6 ms with 70 ppm precision (in the 2 kO to 110 kO range). Two Types of Temperature Correction, Correction by Pt sensor or infrared thermometer. Equipped with EXT I/O, GP-IB and RS-232C interfaces, Easily integrates into automated production lines
The CLIP TYPE LEAD 9287-10 is bundled. As far as concerned other test fixtures, purchase the appropriate lead option for your application separately.
Basic specifications (Accuracy guaranteed for 1 year)
Resistance range : 20 mO range (Max. 20.0000 mO, 0.1 µO resolution) to 100 MO range (Max. 110.000 MO, 1 kO resolution), 11 steps
[at Low Power ON] 2 O range (Max. 2.00000 O, 10 µO resolution) to 2 kO range (Max. 2.00000 kO, 10 mO resolution), 4 steps
Measurement accuracy (SLOW 2) : ±0.007 % rdg. ±0.0015 % f.s. dgt. (at 2 k/ 20 kO range)
[at Low Power ON] ±0.011 % rdg. ±0.01 % f.s. (at 2 to 2 kO range)
Testing current : 100 nA to 1 A DC, [at Low Power ON] 10 µA to 10 mA DC
Open-terminal voltage : 5 V DC (at 20/ 200mO range), 2.6 V DC (at 2 to 20kO range), 13 V DC (at 1M to 100MO range), [at Low Power ON] 60 mV DC
Sampling rate
Delay = 0ms, TC off, Statistics operation off, Offset voltage revise off
FAST MIDIUM SLOW 1 SLOW 2
50Hz 0.6ms ± 0.3ms 21ms ± 1ms 155ms ± 5ms 455ms ± 10ms
60Hz 0.6ms ± 0.3ms 17ms ± 1ms 149ms ± 5ms 449ms ± 10ms
Response time (at truely resistance) : the above time plus approx. 1 ms
(different at setting condition, or testing target)
Other functions : Temperature measurement (-10.0 to 99.9°C), Revised resistance with temperature, Revised temperature with inductor resistance, Comparator, Settings save & load (Max. 30 settings), Classification measurement (BIN function), Statistics operation, Offset voltage revise, Delay, Average function
Interface : RS-232C and GP-IB (standard)
External I/O : [Output] BCD, End of measurement, Comparator result, BIN result, NG, with open collector output
[Input] Select setting, Trigger, Calibration, Zero-adjust, Print, with C-MOS level
Power supply, consumption : 100 to 240 V AC, 50/60 Hz, 30 VA max.
Dimensions, mass : 215 mm (8.46 in) W × 80 mm (3.15 in) H × 295 mm (11.61 in) D, 2.6 kg (91.7 oz)
Supplied accessories : CLIP TYPE LEAD 9287-10 × 1, TEMPERATURE PROBE 9451 × 1, , Power cord × 1, EXT I/O Male connector × 1, Operation manual × 1